A pseudovirus-based neutralizing antibody assay for measuring neutralizing antibody against rabies virus.
Li X, Hu P, Chen Y, +7 more, Sun J.
AI summary — for doctors, generated from the abstract only
A pseudovirus-based assay measured rabies neutralizing antibodies with good precision and strong agreement with the RFFIT gold standard, without needing high-containment facilities.
What was studied
Researchers developed a pseudovirus (pXN2-RABV-G) neutralizing antibody assay for rabies, characterized its analytical performance, and compared it with commercial ELISA and RFFIT using human sera, including 103 vaccine recipients from 2019 to 2023.
Key findings
- The assay showed robustness across culture media and cell passages, acceptable precision, and stability (all P > 0.05).
- Linearity spanned 0.09 to 93.73 IU/mL (R² > 0.95), with a limit of detection of 2.63 IU/mL (95% CI 1.81-3.84) and coefficient of variation under 20%.
- Agreement was substantial with ELISA (kappa 0.74) and almost perfect with RFFIT (kappa 0.82), with strong linear correlation (R² > 0.90).
- Neutralizing antibody titers correlated significantly with vaccine dose; specific statistics not fully captured in the abstract.
Clinical relevance
This assay validation suggests a more accessible way to monitor post-vaccination rabies immunity, but it is a laboratory method study rather than evidence about clinical protection.
Who should care
Limitations
Single-laboratory development and validation with a modest serum sample of 103 recipients, and some reported associations are truncated in the abstract.
A safer, simpler rabies antibody test could make post-vaccination monitoring more widely feasible.
This is an AI-generated summary of the published abstract, not a clinical recommendation. It may not reflect the full text or subsequent updates — verify against the original source before acting on it.
Abstract
To establish a laboratory detection assay for a human neutralizing antibody (NAb) against rabies virus and to evaluate its efficiency in human serum samples. A pseudovirus pXN2‑RABV‑G harboring the RABV G gene was applied. The precision, robustness, linearity, stability, inclusivity, and limit of detection of the assay were evaluated. The consistency and correlation of the assay with ELISA and rapid fluorescent focus inhibition test (RFFIT) were analyzed. A pseudovirus-based neutralizing antibody (pVNA) detection system based on the pXN2‑RABV‑G was successfully developed and showed broad robustness (culture media: χ2 = 4.00, P > 0.05, cell passages: χ2 = 0.25, P > 0.05), precision (independent detection: P > 0.05), lower/upper limit of linearity (0.09 to 93.73 IU/mL, R2 > 0.95), and stability (P > 0.05). The limit of detection (LOD) was 2.63 IU/mL (95% CI: 1.81-3.84), expressed as the lowest human sera antibody level reliably distinguishable from background signals, and the coefficient of variation was less than 20%. Compared with the commercial ELISA and the gold standard RFFIT assay, the kappa values were 0.74 and 0.82, respectively, and the results were highly consistent. Linear correlation analysis revealed a strong correlation between these variables (R2 > 0.90). A total of 103 recipients from 2019 to 2023 were included in the serological testing and the following analysis. NAb titers were significantly correlated with vaccine dose (P P < 0.001). The geometric mean titer (GMT) of the 5-dose vaccination group was greater than that of the 1-dose, 2-dose, 3-dose, and 4-dose groups (3.15 vs 3.19 vs 6.18 vs 12.21 vs 13.40 IU/mL). The established pVNA assay for human rabies NAb detection has a detection capability similar to that of the gold standard RFFIT, but it has the advantages of reduced biosafety requirements, laboratory activity, and time, which highlight its safety, high efficiency, and feasibility for screening and monitoring rabies virus antibodies after vaccination.
Importance
Rabies is a lethal disease with no effective clinical treatment. Vaccination is the only way to prevent it routinely, which makes post-vaccination antibody testing essential. The current standard test requires a highly equipped laboratory, biosafety measures, and is time-consuming. This study develops a safe, efficient test for detecting rabies neutralizing antibodies using pseudovirus technology. It is comparable to the gold standard test in accuracy but easier to implement in routine labs, which enables broader use in screening and monitoring antibodies post-vaccination. This advantage addresses a key gap in rabies prevention, reducing the uncertainty for individuals with immunodeficiency who are nonresponsive to vaccination and at risk of rabies exposure.
DOI
10.1128/jcm.00877-26
PMID
42626959
PMCID
PMC13556208
Research metadata (titles, abstracts, authors, publication dates) is sourced from Europe PMC, which aggregates content from PubMed/MEDLINE, PubMed Central, and other providers under the U.S. National Library of Medicine (NLM/NCBI). NLM does not claim copyright on abstracts; the original journal publishers or authors may hold rights to the full articles. AI-generated summaries are derived only from the abstract, never the full text, and are always labeled as AI-generated.