Back to research feed

Journal of orthopaedic research : official publication of the Orthopaedic Research Society

August 1, 2026 Β· via Europe PMC

Fractures Around Collared and Collarless Hip Stems Initiate in High Strain Regions: A Combined Cadaveric Experiment and Finite Element Analysis.

Helbock RT, Zigan C, Hughes A, +7 more, Quevedo Gonzalez FJ.

research-articleAbstract only β€” full text under license

AI summary β€” for doctors, generated from the abstract only

In cadaveric femurs, collared hip stems withstood higher fracture loads, with benefit tied to initial collar-calcar contact, and fractures began in high-strain regions.

What was studied

A combined cadaveric and finite element study of six paired female femurs implanted with collared or collarless stems and loaded to failure in simulated stumbling.

Key findings

  • Load to fracture was greater for collared stems
  • Fracture load inversely correlated with collar-calcar gap (r=-0.80, p=.055)
  • Experimental fracture location matched the first yielding cortical element in five cases
  • FE-predicted fracture onset load moderately correlated with experimental load (R2=0.43)

Clinical relevance

Suggests achieving calcar contact may be important for collared stems; small cadaveric sample.

Who should care

Orthopaedics

Limitations

Small cadaveric sample of six pairs.

A collar helps protect against periprosthetic fracture mainly when it actually contacts the calcar.

This is an AI-generated summary of the published abstract, not a clinical recommendation. It may not reflect the full text or subsequent updates β€” verify against the original source before acting on it.

Abstract

Periprosthetic femoral fracture (PFF) is a common early complication after primary total hip arthroplasty. Collared stems reduce but do not eliminate the incidence of PFF, and exhibit variable biomechanical effectiveness, emphasizing our limited understanding of the local bone-implant interaction mechanics. Our goal was to elucidate the relationship between the local strains at the bone-implant interface and the experimental fracture patterns and loads for collared and collarless stems. Six pairs of women cadaveric femurs implanted with a collarless or collared stem were loaded to failure under simulated stumbling to determine the location of PFF and the load to fracture, which we related to the collar-calcar separation. Corresponding specimen-specific FE models were developed to determine the strain at the bone-implant interface and to predict the fracture onset load and location of fracture, which we related to the experimental location of PFF and the load-to-fracture. Load to fracture was greater for collared stems and was inversely correlated with the collar-calcar separation at the time of implantation (r = -0.80, p = 0.055). Fractures occurred in areas of high strain. The experimental fracture location coincided with the first yielding cortical element in five cases. The load at which the first cortical element yielded (i.e., onset of fracture was moderately correlated with the experimental fracture load (R2 = 0.43, RMSE = 1231 N). Our results emphasize the importance of initial calcar contact to realize the benefits of the collar. Localized load transfer was a precursor of macroscopic fracture, which initiated in areas of high tensile strain.

Periprosthetic femoral fracture (PFF) is a common early complication after primary total hip arthroplasty. Collared stems reduce but do not eliminate the incidence of PFF, and exhibit variable biomechanical effectiveness, emphasizing our limited understanding of the local bone-implant interaction mechanics. Our goal was to elucidate the relationship between the local strains at the bone-implant interface and the experimental fracture patterns and loads for collared and collarless stems. Six pairs of women cadaveric femurs implanted with a collarless or collared stem were loaded to failure under simulated stumbling to determine the location of PFF and the load to fracture, which we related to the collar-calcar separation. Corresponding specimen-specific FE models were developed to determine the strain at the bone-implant interface and to predict the fracture onset load and location of fracture, which we related to the experimental location of PFF and the load-to-fracture. Load to fracture was greater for collared stems and was inversely correlated with the collar-calcar separation at the time of implantation (r = -0.80, p = 0.055). Fractures occurred in areas of high strain. The experimental fracture location coincided with the first yielding cortical element in five cases. The load at which the first cortical element yielded (i.e., onset of fracture was moderately correlated with the experimental fracture load (R2 = 0.43, RMSE = 1231 N). Our results emphasize the importance of initial calcar contact to realize the benefits of the collar. Localized load transfer was a precursor of macroscopic fracture, which initiated in areas of high tensile strain.

DOI

10.1002/jor.70267

PMID

42609042

PMCID

PMC13482135

Research metadata (titles, abstracts, authors, publication dates) is sourced from Europe PMC, which aggregates content from PubMed/MEDLINE, PubMed Central, and other providers under the U.S. National Library of Medicine (NLM/NCBI). NLM does not claim copyright on abstracts; the original journal publishers or authors may hold rights to the full articles. AI-generated summaries are derived only from the abstract, never the full text, and are always labeled as AI-generated.