Multimodal comparison of 19 single-use flexible ureteroscopes: bench testing, operative video assessment, and surgeon ratings.
Demirkıran ED, Menozzi M, Ventimiglia E, Pini G, Bozzini G, Perri D.
AI summary — for doctors, generated from the abstract only
Single-use flexible ureteroscopes varied markedly across 19 models, and bench metrics did not clearly predict surgeon-reported usability.
What was studied
Bench testing and 57 clinical procedures with 19 devices, rated by surgeon and blinded reviewers.
Key findings
- Resolution ranged 0.447-0.826 lp/mm and field of view 75-110 degrees.
- Deflection fell to 134 degrees with accessories.
- No bench metric associations survived multiple-comparison correction.
Clinical relevance
Informs device selection.
Who should care
Limitations
One surgeon and three procedures per device.
Weigh multiple properties.
This is an AI-generated summary of the published abstract, not a clinical recommendation. It may not reflect the full text or subsequent updates — verify against the original source before acting on it.
Abstract
Purpose
Single-use flexible ureteroscopes (fURSs) differ substantially in optical and mechanical properties, yet comparative evidence remains limited. We performed a multimodal in vitro and clinical assessment of 19 commercially available single-use fURSs to quantify performance variability and examine which device characteristics are associated with surgeon-reported usability.
Methods
Each of 19 models was bench-tested for field of view, laser-fiber safety distance, irrigation flow, deflection, image resolution, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and color discrimination. One surgeon used each device in three separate patients (57 procedures total) and completed a standardized questionnaire after every case; three blinded reviewers scored anonymized intraoperative video. As a secondary analysis, device-level bench metrics were related to patient-level outcomes using a permutation test with correction for multiple comparisons.
Results
All domains showed marked variability (resolution 0.447-0.826 lp/mm; baseline deflection 219-312°, lowest 134° with accessories; empty-channel irrigation 27.9-39.0 mL/min; field of view 75-110°). In secondary, exploratory analysis, no bench metric showed an association with surgeon-reported or video-based outcomes surviving correction for multiple comparisons (54 tests, minimum q = 0.20).
Conclusion
Contemporary single-use fURSs show substantial heterogeneity, and no single device excelled across every domain. Device selection should weigh several performance characteristics against the intended clinical scenario; whether bench performance predicts surgeon-reported usability remains unresolved and warrants larger, adequately powered studies.
DOI
10.1007/s00345-026-06815-1
PMID
42803974
Research metadata (titles, abstracts, authors, publication dates) is sourced from Europe PMC, which aggregates content from PubMed/MEDLINE, PubMed Central, and other providers under the U.S. National Library of Medicine (NLM/NCBI). NLM does not claim copyright on abstracts; the original journal publishers or authors may hold rights to the full articles. AI-generated summaries are derived only from the abstract, never the full text, and are always labeled as AI-generated.