Back to research feed

Journal of neurosurgery

October 9, 2026 · via Europe PMC

Magnetically guided microrobot via minimally invasive percutaneous cisternal access: a first preclinical evaluation.

Mereuta OM, Ognard J, Loudon WG, +14 more, Kadirvel R

Journal ArticleAbstract only — full text under license

AI summary — for doctors, generated from the abstract only

In sheep, a magnet-steered microrobot could be placed through the cisterna magna, moved around and pulled back out without major harm over 28 days.

What was studied

A GLP-compliant large-animal study in 7 sheep (3 treated, 2 controls, 2 alternates) testing percutaneous cisternal access, magnetic navigation, dwell and retrieval of a small coated microrobot using an image-guided electromagnetic platform. The long-term aim is Dandy-Walker cyst fenestration.

Key findings

  • All 3 treated animals completed the full workflow, with first-attempt cannulation and single-maneuver retrieval.
  • No neurological deficits over 28 days; one transient leukocytosis resolved and CSF was normal.
  • Necropsy showed dural fibrosis and thin adhesions in 2 of 3, and small superficial vermian gliosis (2 mm or less) in 2 of 3, without hemorrhage or necrosis.

Clinical relevance

This is an early step toward minimally invasive CSF-space interventions that avoid crossing brain tissue, which could matter for pediatric posterior fossa cystic disease. It is not yet applicable to patients.

Who should care

Neurosurgery

Limitations

Only 3 treated animals, descriptive outcomes, a 28-day follow-up and no efficacy testing limit what can be concluded about safety.

Nothing changes in practice now; treat this as proof of feasibility awaiting longer and larger preclinical work.

This is an AI-generated summary of the published abstract, not a clinical recommendation. It may not reflect the full text or subsequent updates — verify against the original source before acting on it.

Abstract

OBJECTIVE

Magnetically guided microrobots could enable minimally invasive intracranial interventions by navigating CSF spaces without traversing brain parenchyma. However, clinical translation requires in vivo validation of delivery, control, retrieval, and short-term safety under realistic workflow conditions. The aim of this study was to assess the feasibility and initial safety of percutaneous cisterna magna access, microrobot navigation, and retrieval using an image-integrated electromagnetic coil platform in a large-animal model, targeting future application to Dandy-Walker malformation cyst fenestration.

METHODS

In this Good Laboratory Practice-compliant ovine study, 7 male sheep were allocated to an MRI-only control group (n = 1), flexion control group (n = 1), and treatment group (n = 3; percutaneous cisternal access, deployment of a 3.5 × 7-mm Parylene-C-coated bladed microrobot with a neodymium core, magnetic navigation, scripted dwell, and retrieval), with 2 animals designated as alternates. Preoperative MRI was used for planning, with intraoperative cone-beam CT fusion and biplanar fluoroscopy. The follow-up period was 28 days and included daily observation, weekly neurological examination, and analysis of hematology/chemistry, intraoperative CSF, and terminal necropsy with histopathology. Outcomes were descriptive.

RESULTS

The full workflow was completed in all 3 treated animals without conversion or rescue. Cannulation succeeded on the first attempt; continuous fluoroscopic visibility and stable magnetic control were maintained; and retrieval succeeded in a single maneuver. Postoperative recovery was uneventful, with weight gain and no neurological deficits over 28 days of follow-up. Laboratory values were within reference ranges except for 1 case of transient leukocytosis (day 0) that resolved. CSF was acellular/normal. Necropsy showed localized procedure-site changes only (dural fibrosis, thin adhesions in 2 of 3 treated animals). Histology revealed focal superficial vermian gliosis ≤ 2 mm in 2 of 3 treated animals without hemorrhage, necrosis, or off-target injury, and controls were within normal limits (1 case of minimal incidental axonal change in the flexion control).

CONCLUSIONS

Percutaneous cisternal delivery and retrieval of an untethered microrobot were feasible and tolerated in sheep, with localized tissue responses at 28 days. These findings support further studies focused on longer follow-up, workflow optimization, and disease-specific efficacy.

DOI

10.3171/2026.4.JNS252541

PMID

42854255

Research metadata (titles, abstracts, authors, publication dates) is sourced from Europe PMC, which aggregates content from PubMed/MEDLINE, PubMed Central, and other providers under the U.S. National Library of Medicine (NLM/NCBI). NLM does not claim copyright on abstracts; the original journal publishers or authors may hold rights to the full articles. AI-generated summaries are derived only from the abstract, never the full text, and are always labeled as AI-generated.