Anesthesia and analgesia
October 14, 2026 · via Europe PMC
Characterization of the Bispectral Index in Healthy Volunteers Receiving Inhalation Anesthetics in Combination With Opioids in a Randomized Trial.
Barreto Chang OL, Egan T, Breskin I, Merlo I, MacLeod D
AI summary — for doctors, generated from the abstract only
BIS relates strongly to responsiveness across volatile anesthetics in healthy volunteers, with minimal opioid effect.
What was studied
A randomized trial of 211 enrolled volunteers (143 analyzed) across five regimens.
Key findings
- BIS50 ranged from 67.5 to 76.2; BIS95 49.1 to 63.3
- Desflurane had slightly lower values
- AUC was at least 0.95 across groups
Clinical relevance
Helps interpret BIS during anesthesia.
Who should care
Limitations
Young healthy volunteers without noxious stimulation; limited generalizability.
BIS tracks responsiveness well in this setting.
This is an AI-generated summary of the published abstract, not a clinical recommendation. It may not reflect the full text or subsequent updates — verify against the original source before acting on it.
Abstract
BACKGROUND
Processed electroencephalogram (pEEG) monitors are commonly used in anesthesia practice. The relationship between the pEEG and clinical levels of sedation over anesthetic concentrations with and without opioids is not well understood. The purpose of this trial was to investigate the relationship between Bispectral IndexTM (BISTM), clinical measures of sedation, and inhaled anesthetics across a range of anesthetic concentrations and hypnotic states with and without intravenous opioids.
METHODS
This prospective, randomized trial enrolled 211 healthy volunteers. Participants were randomized to one of five anesthetic regimens: sevoflurane, sevoflurane with remifentanil, sevoflurane with fentanyl, desflurane, or isoflurane. Target anesthetic end-tidal concentrations were increased in a stepwise fashion until a BIS index of 30 to 40 was reached. Opioids were administered by target-controlled infusion and commenced before the inhaled agent began. At each targeted end-tidal concentration, responsiveness was assessed using the Modified Observer's Assessment of Alertness/Sedation (MOAA/S) score and the BIS index was recorded. The MOAA/S score was used to assess participant responsiveness, with a score ≤2 considered unresponsive and a score ≥3 considered responsive. The primary outcome measure was the BIS50 for each anesthetic regimen, representing the BIS index at which 50% of participants were unresponsive. Secondary outcomes included BIS95 (representing the BIS index at which 95% of participants were unresponsive) and evaluation of the BIS index as a predictor of responsiveness. BIS50 and BIS95 were estimated using random intercept logistic regression models, and the predictive performance of BIS indices was assessed by calculating the corresponding area under the receiver operating characteristic curve (AUC).
RESULTS
One hundred and forty-three participants were randomized to a treatment group and completed the procedure with adequate data. Mean age was 26.5 ± 6.7 years (52% male). There were no reported device-related adverse events. Across treatment groups, BIS50 ranged from 67.5 (95% confidence interval [CI], 63.6-71.4) to 76.2 (73.5-78.8), and BIS95 ranged from 49.1 (42.9-55.4) to 63.3 (59.5-67.2). Compared to the sevoflurane and isoflurane groups, the desflurane group exhibited slightly lower BIS50 and BIS95 values. The addition of opioids to sevoflurane had minimal impact on BIS50 and BIS95. The BIS index alone was a strong predictor of the level of responsiveness, with AUCs consistently close to the ideal value of 1 (≥0.95) across all treatment groups. BIS indices were highly correlated with MOAA/S scores.
CONCLUSIONS
These results describe the relationship between BIS levels and responsiveness for commonly used volatile anesthetics, with and without opioids, across a range of anesthetic concentrations. The results of this study should be interpreted in light of the study design which included young, healthy adult volunteers with no sustained, noxious stimulation. These findings may not be generalizable to other populations or clinical situations.
DOI
10.1213/ANE.0000000000008332
PMID
42873975
Research metadata (titles, abstracts, authors, publication dates) is sourced from Europe PMC, which aggregates content from PubMed/MEDLINE, PubMed Central, and other providers under the U.S. National Library of Medicine (NLM/NCBI). NLM does not claim copyright on abstracts; the original journal publishers or authors may hold rights to the full articles. AI-generated summaries are derived only from the abstract, never the full text, and are always labeled as AI-generated.